An evaluation of the Rapid Infusion System.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1626672.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The performance of the Rapid Infusion System was evaluated in the laboratory. Using a conventional mixture of two units of packed red cells, two units of fresh frozen plasma and 500 ml crystalloid, a single line and a driving pressure of 300 mmHg, the highest flow in our study was 970 ml.min-1 (2.8 mm catheter, no stopcock). With a 1.6 mm venous cannula the measured flow was 640 ml.min-1. Additional diluting of the standard 'blood cocktail' did not add much to the performance of the system. When primed with tap water 21 degrees C (12 degrees C respectively), the fluid at the outlet of the system reached a maximum temperature of 37.8 degrees C (37.4 degrees C) after 6 min at a flow of 400 ml.min-1. At flows higher than 1150 ml.min-1 (priming with 12 degrees C tap water: 800 ml.min-1), the system slowed down to flows of 700 to 1000 ml.min-1 in order to maintain an adequate temperature. We conclude, that the Rapid Infusion System is a valuable tool for situations where a rapid but controlled replacement with warmed blood at rates up to at least 1100 ml.min-1 is needed. The use of large bore intravenous catheters and avoiding additional resistors such as standard 3-way stopcocks is highly recommended.
Medical subject headings
- Blood Transfusion
- Hot Temperature