Anterior cervical screw extrusion leading to acute upper airway obstruction: case report.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 16284580.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Case report of late postoperative complication. There have been a number of reports of migration and extrusion of cervical fusion instrumentation. The majority of such cases have a benign outcome. To our knowledge, cervical instrumentation extrusion resulting in prevertebral abscess and acute airway obstruction has not been reported. A 56-year-old man who had undergone a prior C3-C6 anterior cervical decompression and fusion presented to a hospital with dysphagia and acute airway obstruction requiring an emergency tracheostomy. His neck radiograph showed that a C6 screw was missing compared to prior films. Magnetic resonance imaging showed a large prevertebral abscess anterior to C2-C7 causing complete upper airway obstruction. He underwent surgical drainage of the abscess and had a good neurologic recovery. We report a case of acute upper airway obstruction from prevertebral abscess, likely secondary to a loosened anterior cervical screw penetrating the prevertebral soft tissue. In contrast to case reports in the literature involving instrumentation extrusion with a usually benign outcome, our case presented with a life-threatening condition.
Medical subject headings
- Airway Obstruction
- Bone Screws
- Cervical Vertebrae
- Decompression, Surgical
- Spinal Fusion
Anatomy
- cervical spine