Cementation of a metal-inlay polyethylene liner into a stable metal shell in revision total hip arthroplasty.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 16310003.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to evaluate the results of liner cementation into a stable acetabular shell using a metal-inlay polyethylene liner in 39 revision total hip arthroplasties. After an average of 2.8 years, 1 cemented liner dislodged from its metal shell at a postoperative 4 years. In the other 38 hips, mean Harris hip scores improved from 65 preoperatively to 86.9 at the final follow-up. Eighteen patients were rated as having excellent results, 17 as good, and 3 as fair. There were no changes in cup position and no cases of osteolytic lesion progression around the femoral and acetabular components in the last follow-up radiographs. Metal-inlay polyethylene liner cementation into a stable acetabular shell was found to provide an alternative option with short-term excellent results, and it also offers more liner options, the preservation of bone stock, and lower surgical morbidity.