Premature failure of the polyethylene tibial bearing surface of the Interax knee arthroplasty.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 16310004.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Thirty-four patients (37 knees) were treated with the Interax knee arthroplasty. There were 31 female and 3 male patients with a mean age at surgery of 70 years (range, 53-84 years). Premature failure of the polyethylene tibial bearing surface occurred in 9 patients (9 knees). Eight patients of these required a revision operation because of early wearing out of the tibial surface 20 to 38 months after the operation. The failure was due to inadequate implant design in small component and polyethylene gamma irradiated in the air.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
- Polyethylene
- Prosthesis Failure
Anatomy
- knee
- tibia