Rapid two-dimensional dose measurement in brachytherapy using plastic scintillator sheet: linearity, signal-to-noise ratio, and energy response characteristics.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1639641.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Because of the large dose gradients encountered near brachytherapy sources, an efficient, accurate, low-atomic number areal detector, which can record dose at many points simultaneously, is highly desirable. We have developed a prototype of such a system using thin plates of plastic scintillator as detectors. A micro-channel plate (MCP) image intensifier was used to amplify the optical scintillation images produced by radioactive 125I and 137Cs sources in water placed 0.5-5.7 cm distance from the detector. A charge-coupled device (CCD) digital camera was used to acquire 2-D light-intensity distributions from the image intensifier output window. For both isotopes, a small area (2 x 3 mm2) PVT detector yields a CCD net count rate that is linear with respect to absorbed dose rate within +/- 3% out to 5.7 cm distance. Acquisition times range from 1.5-400 sec with a reproducibility of 0.5-5.5%. If a large-area (6 x 20 cm2) PVT detector is used, a four-fold increase in count rate and large deviations from linearity are observed, indicating that neighboring pixels contribute light to the signal through diffusion and scattering in PVT and water. A detailed noise analysis demonstrates that the image intensifier reduces acquisition time 10000-fold, reduces noise relative to signal 200-fold, and reduces amplifier gain noise as well.
Medical subject headings
- Brachytherapy
- Radiometry
- Scintillation Counting