Rapid polarizing field cycling in magnetic resonance imaging.
other
Where this comes from
- Record sourced from PubMed, PMID 16398417.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We describe the electronics for controlling the independently pulsed polarizing coil in a prepolarized magnetic resonance imaging (PMRI) system and demonstrate performance with free induction decay measurements and in vivo imaging experiments. A PMRI scanner retains all the benefits of acquiring MRI data at low field, but with the higher signal of the polarizing field. Rapidly and efficiently ramping the polarizing coil without disturbing the data acquisition is one of the major challenges of PMRI. With our modular hardware design, we successfully ramp the 0.4-T polarizing coil of a wrist-sized PMRI scanner at up to 100 T/s without causing image artifacts or otherwise degrading data acquisition.
Medical subject headings
- Electronics, Medical
- Image Enhancement
- Magnetic Resonance Imaging
- Magnetics