The effect of gel lubrication on cuff leakage of double lumen tubes during thoracic surgery.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 16430565.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
High-volume, low-pressure tracheal cuffs of disposable double lumen tubes may offer limited protection to the dependent lung if fluid leaks through folds in the inflated cuffs. This study was undertaken to determine the incidence of fluid leakage past the tracheal cuff and whether gel lubrication reduces the incidence. Fifty-five patients were randomly assigned to receive a double lumen tube with or without gel lubrication. The dependent lung was intubated. With the patient in the lateral position, methylthionium chloride was administered above the tracheal cuff via a pre-attached catheter. Fibreoptic bronchoscopy was performed to determine if dye had passed the tracheal cuff. Three patients were excluded. Dye leakage was seen in 12/27 and 3/25 patients in the unlubricated and lubricated group, respectively (p = 0.014). Gel lubrication significantly reduces fluid leakage past the tracheal cuff of a double lumen tube and should be considered for all thoracic surgical patients requiring one-lung ventilation.
Medical subject headings
- Intubation, Intratracheal
- Lubrication
- Pneumonia, Aspiration
- Postoperative Complications
- Thoracic Surgical Procedures