Experimental study of aortic anastomosis using a circular stapling device in the porcine model.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 16464620.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to assess the strength (pressure resistance) and histological findings of aortic anastomoses performed using a circular stapling device. A circular stapling device was used for anastomosing a porcine aorta and a Dacron graft. The maximum pressure resistance of the anastomotic site of a porcine aortic specimen and a Dacron graft was examined (n=10). A porcine aorta with Dacron graft was anastomosed to a beating heart, and pressure overload was induced by adrenaline (n=5). Specimens of the anastomotic sites were harvested after 14 days and examined histologically. The maximum pressure resistance of the anastomotic site was 427.3+/-34.4 (375-511) mmHg. No anastomotic sites leaked as a result of pressure overloading at 227.6+/-21.1 (201-260) mmHg. Histologically, good incorporation and cell coverage were observed, and the inner surfaces of the anastomotic sites were smooth and without stenoses. Aortic anastomosis using a circular stapling device is feasible and worthy of further investigation.
Medical subject headings
- Anastomosis, Surgical
- Aorta
- Blood Vessel Prosthesis Implantation
- Surgical Staplers