Heating of non-ferromagnetic total hip endoprostheses during magnetic resonance imaging in an animal model.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 16485460.
- Also identified by DOI 10.3928/01477447-20060201-06.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the heating effects of magnetic resonance imaging (MRI) on non-ferromagnetic total hip endoprostheses, hip endoprostheses were tested in vitro and in an ex vivo animal model. The MRI protocol on a 1.5 T scanner consisted of five sequences with a total duration of 23 min. The maximum heating of prosthetic material was 0.17 K in vitro and 0.90 K ex vivo; the maximum heating of bone surrounding tissue was 0.73 K ex vivo. Slight heating of the prostheses and the surrounding tissue was noted, which should not have any side effects in patients with titanium total hip endoprostheses.