Syndesmosis fixation: analysis of shear stress via axial load on 3.5-mm and 4.5-mm quadricortical syndesmotic screws.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 16513499.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The effect of shear stress on a fixated distal syndesmosis of the ankle was evaluated with a servohydraulic materials-testing machine. Eighteen syndesmoses were fixated in a quadricortical technique using 3.5-mm cortical and 4.5-mm cortical stainless steel screws. A shear stress was applied via an axial load in an attempt to simulate weightbearing. The 4.5-mm quadricortical screws produced a higher yield load and peak load (484.3 +/- 93.8 N and 597.7 +/- 81.4 N) when compared with the 3.5-mm quadricortical syndesmotic screws (412.8 +/- 55 N, P = .033 and 511.2 +/- 64.4 N). These findings suggest that a larger diameter screw provides greater resistance to an applied shear stress at the distal syndesmosis.
Medical subject headings
- Ankle Injuries
- Bone Screws
- Fracture Fixation, Internal
- Fractures, Bone
- Stress, Mechanical
- Weight-Bearing
Anatomy
- ankle