Infusion pump delivers over-dosage of propofol as a result of missing syringe support.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 16551915.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We describe the malfunction of a common drug infusion pump. The syringe saddle was missing and allowed the syringe barrel to contact the pump case, which decreased the outward displacement of the syringe clamp. Then, the infusion pump falsely detected a smaller syringe size and consequently delivered an increased infusion rate and overdose of propofol to the patient. More commonly, an incorrectly mounted syringe may increase the outward displacement of the syringe clamp so that the infusion pump falsely detects a larger syringe size, with resultant less than expected infusion rate.
Medical subject headings
- Drug Delivery Systems
- Infusion Pumps
- Propofol
- Syringes