Comparison of ultrasound and implanted seed marker prostate localization methods: Implications for image-guided radiotherapy.

Scarbrough, Todd J; Golden, Nanialei M; Ting, Joseph Y; Fuller, Clifton D; Wong, Adrian; Kupelian, Patrick A; Thomas, Charles R · Int J Radiat Oncol Biol Phys · 2006

retrospective_cohort · Level III

Where this comes from

Abstract

To analyze two methods of image-guided radiotherapy (IGRT) for external beam radiotherapy of prostate cancer. The prostate was localized by ultrasound (US) in lateral (left/right), vertical (anteroposterior), and longitudinal (superior/inferior) dimensions and then by fiducial seed marker (SM) kV X-ray. Assuming initial setup to skin marks as the origin, the mean suggested shifts (for all dimensions) were hypothesized to be similar and within 1 mm of the origin. The three-dimensional distance discrepancy between suggested SM and US shift points was calculated. We hypothesized a mean discrepancy >5 mm as clinically significant. From 40 patients, 1019 US/SM measurements were obtained. Lateral, vertical, and longitudinal dimensional comparisons reveal statistically significant differences in mean shifts (p < 0.0001 for all). US dimensional shifts reveal significantly greater variability. The US three-dimensional vector is greater and more variable than the SM vector (p < 0.0001). The mean US/SM three-dimensional distance discrepancy is 8.8 mm (significantly >5 mm, p < 0.0001). Ultrasound and SM methods suggest different shifts. US data reveal greater systematic/random error vs. SM data. The US data suggest larger PTV expansion margins (approximately 9 mm) are necessary for US IGRT vs. SM IGRT (approximately 3 mm). The hypotheses that US and SM methods suggest similar shifts and that the mean US/SM three-dimensional distance discrepancy is < or =5 mm are rejected.

Medical subject headings