Histologic analysis of a retrieved hydroxyapatite-coated femoral prosthesis.

Søballe, K; Gotfredsen, K; Brockstedt-Rasmussen, H; Nielsen, P T; Rechnagel, K · Clin Orthop Relat Res · 1991

case_report · Level V

Where this comes from

Abstract

A hydroxyapatite-coated hip hemi-prosthesis was retrieved from a 98-year-old osteoporotic woman 12 weeks after implantation. Histologic analysis revealed bone and fibrous tissue almost evenly distributed around the surface of the implant circumference. Quantitative histologic analysis showed that 48% of the hydroxyapatite surface was covered by bone. Fibrous tissue covered 30% of the prosthetic surface, and 20% of the surface had no tissue coverage. Scanning electron microscopy showed direct contact without any clear boundary between the newly formed bone and the hydroxyapatite ceramic.