MR evaluation of pulmonary vein diameter reduction after radiofrequency catheter ablation of atrial fibrillation.

Anselme, Frédéric; Gahide, Gérald; Savouré, Arnaud; Gerbaud, Edouard; Mabru, Mikael; Cribier, Alain; Dacher, Jean-Nicolas · Eur Radiol · 2006

case_series · Level IV

Where this comes from

Abstract

Fifty consecutive patients aged 52+/-12 years suffering from drug refractory atrial fibrillation (AF) underwent baseline and post-ablation MR angiography (MRA) at a mean follow-up of 4+/-3.5 months. Pulmonary vein (PV) disconnection was performed with a maximum energy delivery of 30 W. MRA allowed a two-plane measurement of each PV ostium. After ablation, no significant stenosis was observed, and only 1/194 (0.5%) and 3/194 (2%) PVs had a diameter reduction of 31-40% in the coronal and axial planes, respectively. There was a significant overall post-procedural PV narrowing of 4.9% in the coronal plane and 6.5% in the axial plane (P=ns between both planes). MRA is an efficient technique that can be used in pre- and postoperative evaluation of AF patients. Using a maximal power delivery limited to 30 W, no significant PV stenosis was observed at mid-term follow-up. Late PV anatomical assessment is needed to confirm these results on long-term follow-up.