Transjugular liver biopsy: prospective evaluation of the angle formed between the hepatic veins and the vena cava main axis and modification of a semi-automated biopsy device in cases of an unfavorable angle.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 16683116.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In cases of transjugular liver biopsies, the venous angle formed between the chosen hepatic vein and the vena cava main axis in a frontal plane can be large, leading to technical difficulties. In a prospective study including 139 consecutive patients who underwent transjugular liver biopsy using the Quick-Core biopsy set, the mean venous angle was equal to 49.6 degrees. For 21.1% of the patients, two attempts at hepatic venous catheterization failed because the venous angle was too large, with a mean of 69.7 degrees. In all of these patients, manual reshaping of the distal curvature of the stiffening metallic cannula, by forming a new mean angle equal to 48 degrees , allowed successful completion of the procedure in less than 10 min.
Medical subject headings
- Biopsy, Needle
- Hepatic Veins
- Liver
- Vena Cava, Inferior