Evaluation of MRI RF probes utilizing infrared sensors.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 16686420.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The magnetic resonance imaging (MRI) coil's radio frequency (RF) field distribution has a strong effect on image quality as well as specific absorption rate. In this paper, a method of probing a coil's RF field distribution over any unoccupied region of the coil is presented. This technique is based on the use of infrared sensing. The proposed method was implemented and tested on a high field RF volume coil operating at 340 MHz. Very good agreement was achieved between the infrared measurements and numerical data obtained utilizing an in-house three-dimensional finite-difference time-domain package. The results demonstrate that the proposed technique is practical, robust, and efficient in making accurate measurements of the electric field distributions in loaded and unloaded MRI coils.
Medical subject headings
- Image Enhancement
- Image Interpretation, Computer-Assisted
- Magnetic Resonance Imaging
- Radiometry
- Spectrophotometry, Infrared
- Thermography
- Transducers