MR liver imaging and cholangiography in the presence of surgical metallic clips at 1.5 and 3 Tesla.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 16703314.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate whether clips from prior cholecystectomy impair image quality during magnetic resonance cholangiography (MRC) at 3 Tesla (T) compared with 1.5 T, surgical clips were embedded in a gel phantom and positioned at predefined distances from a fluid-filled tube designed to simulate the bile duct. The maximum clip distance was noted where susceptibility artifacts obscured the fluid-filled tube at 1.5 T and 3 T. Susceptibility artifact size was calculated for each sequence within each magnet class. In vivo analysis included 42 patients postcholecystectomy who underwent MRC at either 1.5 T or 3 T. In vitro, mean area of susceptibility artifacts was 104 mm2 on 3-T and 75 mm2 on 1.5-T MR imaging (MRI). While surgical clips within a 2-mm range impaired visualization of the fluid-filled tube on 1.5-T MRI, this range increased to 4 mm on 3-T MRI. In vivo, MRC image quality was impaired by susceptibility artifacts in three of 21 cases at 3 T and in two of 21 cases at 1.5 T. Overall, biliary pseudo-obstructions due to susceptibility artifacts from cholecystectomy surgical clips were not substantially more common on 3-T MRC in clinical practice, and patients with a history of prior cholecystectomy should not be excluded from a 3-T MRC.
Medical subject headings
- Artifacts
- Bile Duct Diseases
- Magnetic Resonance Imaging
- Surgical Instruments