A comparison of CT- and ultrasound-based imaging to localize the prostate for external beam radiotherapy.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 16751060.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study assesses the accuracy of NOMOS B-mode acquisition and targeting system (BAT) compared with computed tomography (CT) in localizing the prostate. Twenty-six patients were CT scanned, and the prostate was localized by 3 observers using the BAT system. The BAT couch shift measurements were compared with the CT localization. Six of the patients had gold markers present in the prostate, and the prostate movement determined by BAT was compared with the movement determined by the gold markers. Using the BAT system, the 3 observers determined the prostate position to be a mean of 1-5 mm over all directions with respect to the CT. The proportion of readings with a difference >3 mm between the observers was in the range of 25% to 44%. The prostate movement based on gold markers was an average of 3-5 mm different from that measured by BAT. The literature assessing the accuracy and reproducibility on BAT is summarized and compared with our findings. We have found that there are systematic differences between the BAT-defined prostate position compared with that estimated on CT using gold grain marker seeds.
Medical subject headings
- Prostatic Neoplasms