Automatic adjustment of pressure support by a computer-driven knowledge-based system during noninvasive ventilation: a feasibility study.

Battisti, Anne; Roeseler, Jean; Tassaux, Didier; Jolliet, Philippe · Intensive Care Med · 2006

prospective_cohort · Level II

Where this comes from

Abstract

To evaluate the feasibility of using a knowledge-based system designed to automatically titrate pressure support (PS) to maintain the patient in a "respiratory comfort zone" during noninvasive ventilation (NIV) in patients with acute respiratory failure. Prospective crossover interventional study in an intensive care unit of a university hospital. Twenty patients. After initial NIV setting and startup in conventional PS by the chest physiotherapist NIV was continued for 45 min with the automated PS activated. During automated PS minute-volume was maintained constant while respiratory rate decreased significantly from its pre-NIV value (20+/-3 vs. 25+/-3 bpm). There was a trend towards a progressive lowering of dyspnea. In hypercapnic patients PaCO(2) decreased significantly from 61+/-9 to 51+/-2 mmHg, and pH increased significantly from 7.31+/-0.05 to 7.35+/-0.03. Automated PS was well tolerated. Two system malfunctions occurred prompting physiotherapist intervention. The results of this feasibility study suggest that the system can be used during NIV in patients with acute respiratory failure. Further studies should now determine whether it can improve patient-ventilator interaction and reduce caregiver workload.