Model-based RSA of a femoral hip stem using surface and geometrical shape models.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 16826102.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Roentgen stereophotogrammetry (RSA) is a highly accurate three-dimensional measuring technique for assessing micromotion of orthopaedic implants. A drawback is that markers have to be attached to the implant. Model-based techniques have been developed to prevent using special marked implants. We compared two model-based RSA methods with standard marker-based RSA techniques. The first model-based RSA method used surface models, and the second method used elementary geometrical shape (EGS) models. We used a commercially available stem to perform experiments with a phantom as well as reanalysis of patient RSA radiographs. The data from the phantom experiment indicated the accuracy and precision of the elementary geometrical shape model-based RSA method is equal to marker-based RSA. For model-based RSA using surface models, the accuracy is equal to the accuracy of marker-based RSA, but its precision is worse. We found no difference in accuracy and precision between the two model-based RSA techniques in clinical data. For this particular hip stem, EGS model-based RSA is a good alternative for marker-based RSA.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Femur
- Hip Prosthesis
- Models, Theoretical
Anatomy
- femur
- hip