A versatile phantom for quality assurance in the UK Medical Research Council (MRC) RT01 trial (ISRCTN47772397) in conformal radiotherapy for prostate cancer.

Moore, A Rollo; Warrington, A P Jim; Aird, Edwin G A; Bidmead, A Margaret; Dearnaley, David P · Radiother Oncol · 2006

other · Level V

Where this comes from

Abstract

Within the UK RT01 trial the MRC also funded a quality assurance (QA) programme. This included a planning and dosimetry audit at participating centres using a purpose built phantom. Geometrical setup was visually assessed via field shaping around the phantom GTV (to within the order of 1 mm). Within the phantom, ion chamber positional uncertainties were estimated as 0.6 mm (95% CL, k=2). This was the basis for ion chamber measurements in a variety of dose gradients around the PTV closely simulating a patient case. The design provides a representative but reproducible system for QA in the prostate radiotherapy process, from CT scan to treatment. Setup errors are not eliminated, but minimised and estimated.