Disruption of the vascular prosthesis caused by aortic calcification after replacement of the thoracoabdominal aortic aneurysm.

Tanaka, Hiroshi; Okada, Kenji; Yamashita, Teruo; Kawanishi, Yujiro; Matsumori, Masamiti; Okita, Yutaka · Ann Thorac Surg · 2006

case_report · Level V

Where this comes from

Abstract

A 52-year-old man underwent a replacement of the thoracoabdominal aorta. The aorta was severely calcified, and was replaced by a 24-mm woven Dacron (Vascutek, Renfrewshire, Scotland) graft wrapped with the calcificated aneurysmal wall. His postoperative course was uneventful; however, he collapsed on the 18th postoperative day. He underwent an emergent thoracotomy and the wrapped aneurysmal wall was taken down. The prosthesis graft had a 1-mm disruption in the middle portion, which did not relate to the anastomoses. Experimental study ex vivo showed that disruption of the prosthesis could have occurred after a 3-week pulsatile force caused by a seashell simulating aortic calcification.