Impact of electrode insertion depth on intracochlear trauma.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 16949967.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess the effect of cochlear implant (CI) insertion depth and surgical technique on intracochlear trauma. Twenty-one fresh human temporal bones were implanted with CI electrodes and underwent histologic processing and evaluation. Specimens were grouped into 3 categories: 1) soft implantation technique and standard electrode; 2) soft implantation technique and flexible prototype array; 3) forceful implantations and standard electrode. Based on the grading system (1 to 4), 2 numeric values were calculated indicating the overall severity of cochlear damage (trauma indices). Mean trauma index values were 13.8, 36.3, and 59.2 for group 1, 2, and 3, respectively. Differences in cochlear trauma (trauma index) were nonsignificant between specimens in groups 1 and 2 but were significant between groups 1 and 3. This study gives evidence that intracochlear trauma increases with deep insertions. Thus, in cases where cochlear integrity might be important, limited insertions should be achieved.
Medical subject headings
- Cochlea
- Cochlear Implantation
- Cochlear Implants
- Intraoperative Complications