Serum metal levels and bearing surfaces in total hip arthroplasty.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 16950061.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study evaluates steady-state serum metal levels in patients with 4 different combinations of fixation modalities, materials, and bearing couples. Forty patients with a minimum of 5 years of follow-up and with well-functioning primary total hip arthroplasty were recruited to have serum metal levels measured. Serum chromium and cobalt levels in the metal-on-metal cohort were significantly higher (P < .05) than the other 3 cohorts. The noncemented ceramic-polyethylene cohort had significantly lower (P < .05) serum chromium levels compared to cemented and noncemented metal-polyethylene cohorts and significantly higher serum titanium levels compared to the cemented metal-polyethylene cohort which had no titanium-containing implants. Debris generated at the metal head/neck modular junction likely accounts for the significantly lower serum chromium concentration in ceramic-polyethylene bearing couples.