Estimation of images and nonrigid deformations in gated emission CT.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 16967799.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this paper, we propose and test a new iterative algorithm to simultaneously estimate the nonrigid motion vector fields and the emission images for a complete cardiac cycle in gated cardiac emission tomography. We model the myocardium as an elastic material whose motion does not generate large amounts of strain. As a result, our method is based on minimizing an objective function consisting of the negative logarithm of a maximum likelihood image reconstruction term, the standard biomechanical model of strain energy, and an image matching term that ensures a measure of agreement of intensities between frames. Simulations are obtained using data for the four-dimensional (4-D) NCAT phantom. The data models realistic noise levels in a typical gated myocardial perfusion SPECT study. We show that our simultaneous algorithm produces images with improved spatial resolution characteristics and noise properties compared with those obtained from postsmoothed 4-D maximum likelihood methods. The simulations also demonstrate improved motion estimates over motion estimation using independently reconstructed images.
Medical subject headings
- Algorithms
- Gated Blood-Pool Imaging
- Heart
- Image Enhancement
- Image Interpretation, Computer-Assisted
- Imaging, Three-Dimensional
- Tomography, Emission-Computed