Biomechanical comparison of component position and hardware failure in the reverse shoulder prosthesis.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 16990024.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
There has been renewed interest in reverse shoulder arthroplasty for the treatment of glenohumeral arthritis with concomitant rotator cuff deficiency. Failure of the prosthesis at the glenoid attachment site remains a concern. The purpose of this study was to examine glenoid component stability with regard to the angle of implantation. This investigation entailed a biomechanical analysis to evaluate forces and micromotion in glenoid components attached to 12 polyurethane blocks at -15 degrees, 0 degrees, and +15 degrees of superior and inferior tilt. The 15 degrees inferior tilt had the most uniform compressive forces and the least amount of tensile forces and micromotion when compared with the 0 degrees and 15 degrees superiorly tilted baseplate. Our results suggest that implantation with an inferior tilt will reduce the incidence of mechanical failure of the glenoid component in a reverse shoulder prosthesis.
Medical subject headings
- Arthroplasty, Replacement
- Joint Instability
- Joint Prosthesis
- Prosthesis Failure
Anatomy
- shoulder