Detection of small pulmonary nodules in high-field MR at 3 T: evaluation of different pulse sequences using porcine lung explants.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 17013593.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate two MR imaging sequences for the detection of artificial pulmonary nodules inside porcine lung explants. 67 agarose nodules ranging 3-20 mm were injected into ten porcine lungs within a dedicated chest phantom. The signal on T1-weighted images and radiopacity were adjusted by adding 0.125 mmol/l Gd-DTPA and 1.5 g/l of iodine. A T1-weighted three-dimensional gradient-echo (T1-3D-GRE; TR/TE:3.3/1.1 ms, slice:8 mm, flip-angle:10 degrees ) and a T2-weighted half-Fourier fast-spin echo sequence (T2-HF-FSE; TR/TE:2000/66 ms, slice:7 mm, flip-angle:90 degrees ) were applied in axial orientation using a 3-T system (Intera, Philips Medical Systems, Best, The Netherlands), followed by CT (16x0.5 mm) as reference. Nodule sizes and locations were assessed by three blinded observers. In nodules of >10 mm, sensitivity was 100% using 3D-GRE-MRI and 94% using the HF-FSE sequence. For nodules 6-10 mm, the sensitivity of MRI was lower than with CT (3D-GRE:92%; T2-HF-FSE:83%). In lesions smaller than 5 mm, the sensitivity declined to 80% (3D-GRE) and 53% (HF-FSE). Small lesion diameters were overestimated with both sequences, particularly with HF-FSE. This study confirms the feasibility of 3 T-MRI for lung nodule detection. In lesions greater than 5 mm, the sensitivity of the 3D-GRE sequence approximated CT (>90%), while sensitivity and PPV with the HF-FSE sequence were slightly inferior.
Medical subject headings
- Lung Diseases
- Magnetic Resonance Imaging