A dose-response curve for the negative bias pressure of an intrathoracic pressure regulator during CPR.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 17027140.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
An intrathoracic pressure regulator (ITPR) is a device that can be added to the external end of a tracheal tube to create controlled negative airway pressure between positive pressure ventilations. The resulting downward bias of the airway pressure baseline promotes increased venous return and enhanced circulation during CPR and also during hypovolemic shock. In the present study, we exercised a mathematical model of the human cardiopulmonary system, including airways, lungs, a four chambered heart, great vessels, peripheral vascular beds, and the biomechanics of chest compression and recoil, to determine the relationship between systemic perfusion pressure during CPR and the value of baseline negative airway pressure in an ITPR. Perfusion pressure increases approximately 50% as baseline airway pressure falls from zero to -10 cm H2O. Thereafter perfusion pressure plateaus. Negative bias pressures exceeding -10 cm H2O are not needed in ITPR-CPR.
Medical subject headings
- Cardiopulmonary Resuscitation
- Heart Arrest
- Respiration, Artificial