A national field study of quality assessment of CoaguChek point-of-care testing prothrombin time monitors.

Meijer, Piet; Kluft, Cornelis; Poller, Leon; van der Meer, Felix J M; Keown, Michelle; Ibrahim, Saied; van den Besselaar, Anton M H P; Tripodi, Armando et al. · Am J Clin Pathol · 2006

cross_sectional · Level IV

Where this comes from

Abstract

A system for quality assessment (QA) of the CoaguChek (Roche Diagnostics, Mannheim, Germany) point-of-care testing prothrombin time monitor has been developed by the European Concerted Action on Anticoagulation. Hitherto there has not been an adequate rapid method for CoaguChek QA. Sets of 5 certified international normalized ratio (INR) plasma samples were tested on 539 CoaguChek monitors by experienced staff at 9 Netherlands Thrombosis Centers and results compared with certified INR. A 15% or more deviation has been classified as significant deviation. Overall mean and certified INR values were similar, but 20.3% of participants showed a 15% or more deviation from the certified INR on at least 1 of the 5 QA plasma samples. Statistically significant differences in results with different lots of CoaguChek test strips were found. There is need for large scale QA of CoaguChek monitors. The importance of the 5 CoaguChek certified INR QA plasma samples being tested on a single occasion is demonstrated.