Multiple failed external defibrillation attempts during robot-assisted internal mammary harvest for myocardial revascularization.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 17056941.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We describe multiple failed external defibrillation attempts via rescue defibrillation pads for ventricular fibrillation that occurred secondary to direct electrical current transmission through the pericardial sac from electrocautery during robot-assisted internal mammary harvest. Only after resumption of two-lung ventilation and decompression of the iatrogenic pneumothorax was the patient successfully defibrillated. Conditions necessary for robotic intrathoracic surgery may make defibrillation and resuscitation difficult if they become necessary.
Medical subject headings
- Defibrillators
- Mammary Arteries
- Myocardial Revascularization
- Robotics