Preclinical biocompatibility assessment of the EVAHEART ventricular assist device: coating comparison and platelet activation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 17109415.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Thromboembolism and bleeding remain significant complications of ventricular assist device (VAD) support. Increasing the amount of biocompatibility data collected during preclinical studies can provide additional criteria to evaluate device refinements, while design changes may be implemented before entering clinical use. Twenty bovines were implanted with the EVAHEART centrifugal VAD for durations from 30 to 196 days. Titanium alloy pumps were coated with either diamond-like carbon or 2-methoxyethyloylphosphoryl choline (MPC). Activated platelets and platelet microaggregates were quantified by flow cytometry, including two new assays to quantify bovine platelets expressing CD62P and CD63. Temporally, all assays were low preoperatively, then significantly increased following VAD implantation, before declining to a lower, but still elevated level over 2-3 weeks. MPC-coated VADs produced significantly fewer activated platelets after implant trauma effects diminished. Three animals receiving no postoperative anticoagulation had similar amounts of circulating activated platelets and platelet microaggregates as animals receiving warfarin anticoagulation. Two new methods to quantify bovine activated platelets using antibodies to CD62P and CD63 were characterized and applied. These measures, along with previously described assays, were able to differentiate between two biocompatible coatings and assess effects of anticoagulation regimen in VAD preclinical testing.
Medical subject headings
- Blood Platelets
- Coated Materials, Biocompatible
- Heart-Assist Devices
- Materials Testing
- Platelet Activation