Forehead pulse oximetry: Headband use helps alleviate false low readings likely related to venous pulsation artifact.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 17122573.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study investigated whether a tensioning headband that applies up to 20 mmHg pressure over a forehead pulse oximetry sensor could improve arterial hemoglobin oxygen saturation reading accuracy in presence of venous pooling and pulsations at the forehead site. Healthy volunteers were studied breathing room air in supine and various levels of negative incline (Trendelenburg position) using the forehead sensor with the headband adjusted to its maximum and minimum recommended pressure limits. Saturation readings obtained from the forehead sensor with the subjects supine and the headband in place were used as a baseline to compare the effects of negative incline on reading accuracy when using and not using the headband. Occurrences of false low-saturation readings detected by forehead sensors were compared with those from digit sensors. No difference was observed between saturation readings obtained from the forehead sensor in supine and negative incline positions when the headband was applied. Forehead sensor readings obtained while subjects were inclined and the headband was not used were significantly lower (P < 0.05) than the supine readings. There was no statistically significant difference between the digit and forehead sensor in reporting false low-saturation readings when the headband was applied, regardless of body incline. Application of up to 20 mmHg pressure on the forehead pulse oximetry sensor using an elastic tensioning headband significantly reduced reading errors and provided consistent performance when subjects were placed between supine and up to 15 degrees head-down incline (Trendelenburg position).
Medical subject headings
- Forehead
- Oximetry