Comparison of the stability of various internal fixators used in the treatment of osteochondritis dissecans--a mechanical model.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 17205560.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to determine what characteristics of fixation devices used in the treatment of osteochondritis dissecans (OCD) contribute to improved stability to resist shear loading. An OCD model was designed using rigid polyurethane foam. Each specimen consisted of two components, an osteochondral fragment and a corresponding defect. A total of 40 specimens were prepared and assigned to one of four groups: control (no extrinsic stabilizer); two 2-mm-diameter Kirschner wires (K-wires), 40 mm in length; one threaded washer and a 28-mm screw; and one threaded washer and a 38 mm screw. Each specimen was mounted onto an Iosipescu shear test fixture and subjected to shear loads at a pseudo-static displacement rate of 0.075 mm/s. All groups demonstrated some stability; controls were significantly less stable than all other groups. The group with the threaded washer and 38-mm screw demonstrated the greatest stability (p < 0.001), and no difference was noted between the K-wire and 28-mm screw groups. These results suggest that, in this OCD model, friction conferred some intrinsic stability to resist loads in shear. However, stability was improved with the use of long implants that compressed the fragments together.
Medical subject headings
- Internal Fixators
- Joint Instability
- Materials Testing
- Models, Biological
- Osteochondritis Dissecans