Cementless alumina-on-alumina total hip arthroplasty in patients younger than 50 years: a 5-year minimum follow-up study.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 17275631.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Sixty-seven patients (78 hips) who were younger than 50 years underwent cementless total hip arthroplasty using a porous-coated acetabular cup, an alumina liner, a 28-mm alumina head, and a proximally porous-coated femoral stem. One patient (1 hip) died and 2 patients (3 hips) were lost to follow-up. The remaining 64 patients (74 hips) were followed for 5 to 6 years. Their mean age at the index operation was 37 years. The mean Harris hip score was 94 points at the time of final follow-up. Four patients (4 hips) had mild thigh pain. All acetabular and femoral components were bone-ingrown, and neither pelvic nor femoral osteolysis was identified. No fracture of the ceramic liner or head was identified. Wear of the ceramic components was undetectable in 27 hips in which measurement was possible.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
Anatomy
- hip