A mu-rhythm matched filter for continuous control of a brain-computer interface.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 17278584.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A brain-computer interface (BCI) is a system that provides an alternate nonmuscular communication/control channel for individuals with severe neuromuscular disabilities. With proper training, individuals can learn to modulate the amplitude of specific electroencephalographic (EEG) components (e.g., the 8-12 Hz mu rhythm and 18-26 Hz beta rhythm) over the sensorimotor cortex and use them to control a cursor on a computer screen. Conventional spectral techniques for monitoring the continuous amplitude fluctuations fail to capture essential amplitude/phase relationships of the mu and beta rhythms in a compact fashion and, therefore, are suboptimal. By extracting the characteristic mu rhythm for a user, the exact morphology can be characterized and exploited as a matched filter. A simple, parameterized model for the characteristic mu rhythm is proposed and its effectiveness as a matched filter is examined online for a one-dimensional cursor control task. The results suggest that amplitude/phase coupling exists between the mu and beta bands during event-related desynchronization, and that an appropriate matched filter can provide improved performance.
Medical subject headings
- Algorithms
- Cerebral Cortex
- Electroencephalography
- Evoked Potentials
- Imagination
- Pattern Recognition, Automated
- User-Computer Interface