Interlaboratory and intralaboratory variability in pulmonary function testing. A 13-year study using a normal biologic control.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1729118.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to document the precision with which measurements can be made on a normal subject as a "biologic control" when comparisons are made both on an intralaboratory and interlaboratory basis. One individual had a total of about 300 routine pulmonary function measurements in 22 laboratories during a 13-year period. In this study, significant variation was found in interlaboratory testing, while less variation was found in intralaboratory measurements. This technique of periodic testing of normal individuals as biologic controls is useful for documenting reproducibility and precision; evaluation of accuracy will require a check of the physical calibration of the equipment. The testing (on at least a monthly basis) of a normal nonsmoking subject as a biologic control is recommended as an integral component of quality assurance in the pulmonary function laboratory.
Medical subject headings
- Laboratories, Hospital
- Respiratory Function Tests