Imaging of the human torso using cone-beam transmission CT implemented on a rotating gamma camera.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 1730983.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Radionuclide transmission CT generated on a rotating gamma camera can improve SPECT imaging by providing attenuation maps for attenuation compensation and for anatomical correlation. This paper demonstrates the feasibility and high quality of cone-beam transmission CT (CB-CT) of human subjects, in comparison to conventional parallel-ray CT, and evaluates some possible imaging protocols. Two CB-CT implementation modes, with a cone-beam collimator and without any collimator, were evaluated. Three human subjects of different dimensions were imaged. For the two smaller subjects, the CB-CT images were dramatically superior, in terms of noise and resolution, to those obtained with a parallel-ray geometry. The image noise was less by a factor of 6. CB-CT linear attenuation coefficients were found to be in close agreement with published values for various tissues. For the largest subject, image truncation produced a ring artifact at the edge, but inside the artifact, the image quality was still very good. Cone-beam images obtained without any collimator were acceptable, but photon scatter degraded the image contrast.
Medical subject headings
- Liver
- Thorax
- Tomography, Emission-Computed, Single-Photon
- Tomography, X-Ray Computed