New method for describing the performance of cardiac surgery cannulas.
Level V
Where this comes from
- Record sourced from PubMed, PMID 1731669.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cardiac surgery cannulas are characterized by external diameter only, which provides little information about the pressure-flow characteristics of a cannula. A system has been developed to describe pressure-flow characteristics with a single, unitless number, M, which is patterned after a Reynolds friction factor correlation. A cannula with a lower M number has a more favorable pressure-flow relationship. The M number was determined for 16 arterials cannulas ranging in size from 10F to 26F and 27 venous cannulas sized 12F to 36F. Pressure-flow characteristics vary considerably among cannulas from different manufacturers despite having similar French sizes. Clinical decisions regarding choice of cannula can be simplified by using the M number, which gives a more accurate description of the performance characteristics of a cannula than the French size designation.