Diagnostic value of microtympanometry in primary care.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1737151.
- Also identified by PMC identifier 1881018.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the reliability, validity, and feasibility of a new hand held microtympanometer. Comparison of microtympanometry by two independent observations of a general practitioner and a nurse, and against a validated reference instrument. Primary care health centre of a school for the deaf in the United States. 111 schoolchildren receiving a regular check up. Tympanometry with the Grason Stadler 28, classified with a slightly modified Jerger's classification. Interobserver reliability was 0.95 (Cohen's kappa). Results of microtympanometry were highly comparable with results of the reference instrument (likelihood ratio of positive results, 161.2). The microtympanometer could be used in general practice: it is hand held, child friendly, easy to handle, and accurate.
Medical subject headings
- Acoustic Impedance Tests
- Otitis Media with Effusion