Changes in the light intensity of the fiberoptic laryngoscope blade by steam sterilization.
Level III
Where this comes from
- Record sourced from PubMed, PMID 17377105.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
I investigated the effects of steam sterilization on light intensity of reusable fiberoptic laryngoscope blades. Six new reusable fiberoptic Macintosh laryngoscope blades were steam sterilized. The light intensity was measured, the tip of the blades was photographed, and the light intensity of six new plastic disposable Macintosh laryngoscope blades was also measured. The light intensity of reusable blades exceeded that of the disposable blades after 20 sterilizations, but was significantly lower after 80 sterilizations. The photographs showed irregular lighting of the blades, which increased with repeated sterilization. Disposable plastic blades provide more illumination than reusable fiberoptic blades subjected to repeated sterilization.
Medical subject headings
- Fiber Optic Technology
- Laryngoscopes
- Steam
- Sterilization