Axillary and rectal temperature measurements in infants.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 1739325.
- Also identified by PMC identifier 1793545.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Rectal and axillary temperatures were measured during the daytime in 281 infants seen randomly at home and 656 at hospital under 6 months old, using mercury-in-glass thermometers. The normal temperature range derived from the babies at home was 36.7-37.9 degrees C for rectal temperature and 35.6-37.2 degrees C for axillary temperature. Rectal temperature was higher than axillary in 98% of the measurements. The mean (SD) difference between rectal and axillary temperatures was 0.7 (0.5) degrees C, with a range of 3 degrees C. When used in hospital to detect high temperature, axillary temperature had a sensitivity of 73% compared with rectal temperature. This is too insensitive for accurate detection of an infant's high temperature. Rectal temperature measurement is safer than previously suggested: perforation has occurred in less than one in two million measurements. If an infant's temperature needs to be taken, rectal temperature should be used.
Medical subject headings
- Body Temperature