The influence of electrolyte composition on the in vitro charge-injection limits of activated iridium oxide (AIROF) stimulation electrodes.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 17409482.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The effects of ionic conductivity and buffer concentration of electrolytes used for in vitro measurement of the charge-injection limits of activated iridium oxide (AIROF) neural stimulation electrodes have been investigated. Charge-injection limits of AIROF microelectrodes were measured in saline with a range of phosphate buffer concentrations from [PO(4)(3-)] = 0 to [PO(4)(3-)] = 103 mM and ionic conductivities from 2-28 mS cm(-1). The charge-injection limits were insensitive to the buffer concentration, but varied significantly with ionic conductivity. Using 0.4 ms cathodal current pulses at 50 Hz, the charge-injection limit increased from 0.5 mC cm(-2) to 2.1 mC cm(-2) as the conductivity was increased from 2 mS cm(-1) to 28 mS cm(-1). An explanation is proposed in which the observed dependence on ionic conductivity arises from non-uniform reduction and oxidation within the porous AIROF and from uncorrected iR-drops that result in an overestimation of the redox potential during pulsing. Conversely, slow-sweep-rate cyclic voltammograms (CVs) were sensitive to buffer concentration with the potentials of the primary Ir(3+)/Ir(4+) reduction and oxidation reactions shifting approximately 300 mV as the buffer concentration decreased from [PO(4)(3-)] = 103 mM to [PO(4)(3-)] = 0 mM. The CV response was insensitive to ionic conductivity. A comparison of in vitro AIROF charge-injection limits in commonly employed electrolyte models of extracellular fluid revealed a significant dependence on the electrolyte, with more than a factor of 4 difference under some pulsing conditions, emphasizing the need to select an electrolyte model that closely matches the conductivity and ionic composition of the in vivo environment.
Medical subject headings
- Electric Stimulation
- Electrodes, Implanted
- Electrolytes
- Iridium
- Microelectrodes