The isolation mode rejection ratio in bioelectric amplifiers.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1748450.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Galvanic isolation of a patient during a bioelectric recording is necessary to ensure the safety of the patient. In a typical measurement situation high interference voltages may be present across the isolation barrier. In this paper the necessity of a very high isolation mode rejection ratio--the ability of the amplifier to suppress feed-through from voltages across the isolation barrier to the output--is argued and a design of a multichannel amplifier with an isolation mode rejection ratio of 160 dB is described.
Medical subject headings
- Electrodiagnosis