Biomechanical and functional testing of plate fixation devices for proximal phalangeal fractures.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 1748770.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Functional testing in fresh cadaver digits of a dorsally applied mini-H plate, a mini-straight plate, and a laterally applied mini-condylar plate demonstrated that all three significantly reduce (p less than 0.2) simulated active PIP joint flexion. The mini-condylar plate, however, provided the least reduction of any of the three plates. Biomechanical testing of the same three plates on proximal phalanx after osteotomy in apex palmar and apex dorsal direction bending showed all three plates to provide rigidity less than the intact bone.
Medical subject headings
- Bone Plates
- Finger Injuries
- Finger Joint
- Fractures, Bone
Anatomy
- hand