Fixation strength of unicortical versus bicortical C1-C2 transarticular screws.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 17526435.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The internal carotid artery and hypoglossal nerve lie over the anterior aspect of the lateral mass of the atlas and are at risk from bicortical C1-C2 transarticular screws. This has led to the recommendation for unicortical screws if the neurovascular structures are in close proximity to the proposed exit point. No data are available on strength of unicortical versus bicortical C1-C2 transarticular screws. To compare the biomechanical pullout strength of unicortical versus bicortical C1-C2 transarticular screws in a cadaveric model. Biomechanical study. Fifteen cervical spine specimens underwent axial pullout testing. A unicortical C1-C2 transarticular screw was placed on one side with a contralateral bicortical screw. Data were analyzed to reveal any significant differences in strength. Mean pullout strength for the bicortical C1-C2 transarticular screws was 1,048.8 (+/-360.1) N versus 939.2 (+/-360.6) for unicortical screws (p=.22). There was no significant difference in the pullout strength of unicortical and bicortical screws. In cases with satisfactory bone quality, it appears reasonable to use unicortical screws to avoid the risk of neurovascular injury from penetrating the anterior cortex of C1.
Medical subject headings
- Bone Screws
- Cervical Vertebrae
- Materials Testing
- Shear Strength
- Spinal Fusion
Anatomy
- cervical spine