Use of covered stents and endograft as a rescue treatment in a patient with a complex form of recurrent aortic coarctation.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 17543695.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report our experience with one patient with recurrent coarctation and aneurysm degeneration treated endoluminally. A novel technique was used combining balloon-expandable with self-expandable endografts. The balloon-expandable component expanded the area of stenosis and also created a neck to allow the implantation of the self-expandable endograft to exclude the aneurysm. Mismatch of diameters was solved with the creation of a new proximal neck, in which the distal end was progressively expanded to reach the appropriate diameter to implant the self-expandable endograft inside. Expansion of the narrow stenosis and complete exclusion of the aneurysm was achieved without a residual pressure gradient.
Medical subject headings
- Angioplasty, Balloon
- Aortic Aneurysm, Thoracic
- Aortic Coarctation
- Blood Vessel Prosthesis
- Blood Vessel Prosthesis Implantation
- Stents
- Vascular Surgical Procedures