Precision required for dose-escalated treatment of spinal metastases and implications for image-guided radiation therapy (IGRT).
other
Where this comes from
- Record sourced from PubMed, PMID 17561294.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the precision required in dose-escalated IMRT treatment of spinal metastases and paraspinal tumors. In IMRT treatment plans of nine patients with spinal metastases (n=7) and paraspinal tumors (n=2) translational patient positioning errors (0-10mm) and rotational errors (0-7.5 degrees ) were simulated. The dose to the spinal cord (D5(spine)) resulting from these simulations was evaluated and NTCP for spinal cord necrosis was calculated. All patient set-up errors observed during treatment were simulated and the influence on D5(spine) was investigated. To keep the dose distribution to the spinal cord within +/-5% (+/-10%) of the prescribed dose, maximum tolerable errors of 1mm (2mm) in the transversal plane, 4mm (7mm) in superior-inferior direction and maximum rotations of 3.5 degrees (5 degrees ) were calculated on average. The translational and rotational component of clinically observed set-up errors increased D5(spine) by 23+/-14% and 3+/-2% on average, respectively. Steep dose gradients of IMRT planning require very high precision. In selected patients correction of both translational and rotational errors may be beneficial.
Medical subject headings
- Radiotherapy Planning, Computer-Assisted
- Radiotherapy, Conformal
- Spinal Neoplasms