Precision required for dose-escalated treatment of spinal metastases and implications for image-guided radiation therapy (IGRT).

Guckenberger, Matthias; Meyer, Juergen; Wilbert, Juergen; Baier, Kurt; Bratengeier, Klaus; Vordermark, Dirk; Flentje, Michael · Radiother Oncol · 2007

other

Where this comes from

Abstract

To evaluate the precision required in dose-escalated IMRT treatment of spinal metastases and paraspinal tumors. In IMRT treatment plans of nine patients with spinal metastases (n=7) and paraspinal tumors (n=2) translational patient positioning errors (0-10mm) and rotational errors (0-7.5 degrees ) were simulated. The dose to the spinal cord (D5(spine)) resulting from these simulations was evaluated and NTCP for spinal cord necrosis was calculated. All patient set-up errors observed during treatment were simulated and the influence on D5(spine) was investigated. To keep the dose distribution to the spinal cord within +/-5% (+/-10%) of the prescribed dose, maximum tolerable errors of 1mm (2mm) in the transversal plane, 4mm (7mm) in superior-inferior direction and maximum rotations of 3.5 degrees (5 degrees ) were calculated on average. The translational and rotational component of clinically observed set-up errors increased D5(spine) by 23+/-14% and 3+/-2% on average, respectively. Steep dose gradients of IMRT planning require very high precision. In selected patients correction of both translational and rotational errors may be beneficial.