Extensively porous-coated stems for femoral revision: a choice for all seasons.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 17570290.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We reviewed the outcome of 905 femoral revisions performed at our institution between June 1980 and August 2006 using extensively porous-coated, cylindrical stems. A total of 20 femoral rerevisions were identified, including 12 for aseptic loosening, 4 for infection, 3 for stem fracture, and 1 in conjunction with a periprosthetic femur fracture. All stem rerevisions occurred within the first 10 years after surgery. Kaplan-Meier survivorship, using stem rerevision for any reason as an end point, was 97.5% +/- 1.3% (95% confidence intervals) at 5 years, declined to 95.9% +/- 1.9% at 10 years, and remained constant thereafter. With long-term follow-up, extensively porous-coated, cylindrical stems used for femoral revision continue to demonstrate excellent survivorship.
Medical subject headings
- Hip Prosthesis
Anatomy
- femur
- hip