Impact of operator-selected image noise index and reconstruction slice thickness on patient radiation dose in 64-MDCT.

Kanal, Kalpana M; Stewart, Brent K; Kolokythas, Orpheus; Shuman, William P · AJR Am J Roentgenol · 2007

basic_science · Level V

Where this comes from

Abstract

Our objective was to develop a better understanding of the complex interrelationship between image noise, reconstruction slice thickness, and patient radiation dose on a 64-MDCT scanner that uses automated tube current modulation. We reviewed physics theory and performed phantom dose measurements on a 64-MDCT scanner while altering operator-selectable image noise and reconstruction slice thickness. Using phantom dose measurements to adjust theoretic predictions, we constructed both a spreadsheet and a graph that visually display the interrelationships between operator-selected image noise and reconstruction slice thickness and the resulting patient dose. This table and graph may help operators understand the trade-offs when prospectively trying to minimize dose and optimize image noise for selected reconstruction slice thicknesses on this type of 64-MDCT scanner.

Medical subject headings