Impact of operator-selected image noise index and reconstruction slice thickness on patient radiation dose in 64-MDCT.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 17579174.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Our objective was to develop a better understanding of the complex interrelationship between image noise, reconstruction slice thickness, and patient radiation dose on a 64-MDCT scanner that uses automated tube current modulation. We reviewed physics theory and performed phantom dose measurements on a 64-MDCT scanner while altering operator-selectable image noise and reconstruction slice thickness. Using phantom dose measurements to adjust theoretic predictions, we constructed both a spreadsheet and a graph that visually display the interrelationships between operator-selected image noise and reconstruction slice thickness and the resulting patient dose. This table and graph may help operators understand the trade-offs when prospectively trying to minimize dose and optimize image noise for selected reconstruction slice thicknesses on this type of 64-MDCT scanner.
Medical subject headings
- Body Burden
- Models, Biological
- Radiographic Image Enhancement
- Radiographic Image Interpretation, Computer-Assisted
- Radiometry
- Tomography, X-Ray Computed