Safety and outcomes of balloon catheter sinusotomy: a multicenter 24-week analysis in 115 patients.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 17599558.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to further evaluate the safety and effectiveness of balloon catheter devices to dilate obstructed sinus ostia/perform sinusotomy. Through a prospective, multicenter evaluation, safety was assessed by rate of adverse events, patency was determined by endoscopic examination, and sinus symptoms were determined by the Sino-Nasal Outcome Test (SNOT 20). At the conclusion of the 24-week analysis, endoscopy determined that the sinusotomy was patent in 80.5% (247 of 307) sinuses and nonpatent in 1.6% (5 of 307), and could not determine ostial patency status in 17.9% (55 of 307). Of the ostia visualized on endoscopy, 98% were patent (247 of 252), while 2% (5 of 252) were considered nonpatent. SNOT 20 scores showed consistent symptomatic improvement over baseline. Revision treatment was required in 3 sinuses (3 of 307 sinuses, 0.98%) in 3 patients (3 of 109 patients, 2.75%). Balloon catheter technology appears safe and effective in relieving ostial obstruction. Patients were pleased and indicated that they experienced symptomatic improvement.
Medical subject headings
- Catheterization
- Sinusitis